In vivo evaluation of a chip based near infrared sensor for continuous glucose monitoring.

Authors :
Ben Mohammadi, Lhoucine
Klotzbuecher, Thomas
Sigloch, Susanne
Welzel, Knut
Göddel, Michael
Pieber, Thomas R.
Schaupp, Lukas
Source :
Biosensors & Bioelectronics. Mar2014, Vol. 53, p99-104. 6p.
Publication Year :
2014

Abstract

Abstract: In this paper we describe the concept and in vivo results of a minimally invasive, chip-based near infrared (NIR) sensor, combined with microdialysis, for continuous glucose monitoring. The sensor principle is based on difference absorption spectroscopy in selected wavelength bands of the near infrared spectrum (1300nm, 1450nm, and 1550nm) in the 1st overtone band. In vitro measurements revealed a linear relationship between glucose concentration and the integrated difference spectroscopy signal with a coefficient of determination of 99% in the concentration range of 0–400mg/dl. The absolute error in this case is about 5mg/dl, corresponding to a relative error of about 5% for glucose concentrations larger than 50mg/dl and about 12% in the hypoglycemic range (<50mg/dl). In vivo measurements on 10 patients showed that the NIR-CGM sensor data reflects the blood reference values adequately, if a proper calibration and a signal drift correction is applied. The mean MARE (mean absolute relative error) value taken over all patient data is 13.8%. The best achieved MARE value is at 4.8%, whereas the worst lies at 25.8%, leading to a standard deviation of 5.5%. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
09565663
Volume :
53
Database :
Academic Search Index
Journal :
Biosensors & Bioelectronics
Publication Type :
Academic Journal
Accession number :
92652152
Full Text :
https://doi.org/10.1016/j.bios.2013.09.043