Back to Search Start Over

Submicroliter Electrochemistry and Spectroelectrochemistry Using Standard Electrodes and a Polymer Electrolyte Salt Bridge.

Authors :
Flowers, Paul A.
Blake, David A.
Source :
Analytical Chemistry. 3/19/2013, Vol. 85 Issue 6, p3059-3063. 5p.
Publication Year :
2013

Abstract

The development of spectroscopic and electrochemical devices that can accommodate very small samples is of considerable importance to many areas of science and technology. We report here on the design and characteristics of a simple apparatus for the electrochemical and spectroelectrochemical analysis of submicroliter aqueous samples. The device is easily assembled from common laboratory materials and equipment, utilizing a bifurcated fiber-optic probe, standard disk electrodes of millimeter dimensions, and a polymer electrolyte film salt bridge to enable the analysis of nanoliter-scale sample volumes in a thin-layer configuration. Excellent performance has been demonstrated via measurements on aqueous ferricyanide solutions using sample volumes as low as 20 nL. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00032700
Volume :
85
Issue :
6
Database :
Academic Search Index
Journal :
Analytical Chemistry
Publication Type :
Academic Journal
Accession number :
86725842
Full Text :
https://doi.org/10.1021/ac303712v