Back to Search Start Over

Influence of the active area size and read-out method on the timing performance of SiPMs coupled to LYSO scintillators

Authors :
Tarolli, Alessandro
Gola, Alberto
Ferri, Alessandro
Pro, Tiziana
Serra, Nicola
Zorzi, Nicola
Piemonte, Claudio
Source :
Nuclear Instruments & Methods in Physics Research Section A. Feb2013, Vol. 702, p77-79. 3p.
Publication Year :
2013

Abstract

Abstract: In this work we report on the coincidence resolving time performance of SiPMs with different sizes, produced at FBK, coupled to the same LYSO scintillators. The measurements are performed both with and without the differential leading edge discriminator at three different temperatures, 20°C, 0°C and −20°C. The photo-detectors feature an active area of 2×2mm2 and 4×4mm2. The scintillators have a cross-section of 1.8×1.8mm2 and height of 10mm. The measurements show that, once we eliminate the effect of noise on the timing measurements, we obtain similar coincidence resolving times for the two SiPM sizes considered. This means that the SiPM capacitance, at least up to 4×4mm2, is not a limiting factor. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
01689002
Volume :
702
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
85279155
Full Text :
https://doi.org/10.1016/j.nima.2012.09.012