Interfraction rotation of the prostate as evaluated by kilovoltage X-ray fiducial marker imaging in intensity-modulated radiotherapy of localized prostate cancer
- Source :
-
Medical Dosimetry . Winter2012, Vol. 37 Issue 4, p396-400. 5p. - Publication Year :
- 2012
-
Abstract
- Abstract: To quantify the daily rotation of the prostate during a radiotherapy course using stereoscopic kilovoltage (kV) x-ray imaging and intraprostatic fiducials for localization and positioning correction. From 2005 to 2009, radio-opaque fiducial markers were inserted into 38 patients via perineum into the prostate. The ExacTrac/Novalis Body X-ray 6-day image acquisition system (ET/NB; BrainLab AG, Feldkirchen, Germany) was used to determine and correct the target position. During the first period in 10 patients we recorded all rotation errors but used only Y (table) for correction. For the next 28 patients we used for correction all rotational coordinates, i.e., in addition Z (superior-inferior [SI] or roll) and X (left-right [LR] or tilt/pitch) according to the fiducial marker position by use of the Robotic Tilt Module and Varian Exact Couch. Rotation correction was applied above a threshold of 1° displacement. The systematic and random errors were specified. Overall, 993 software-assisted rotational corrections were performed. The interfraction rotation errors of the prostate as assessed from the radiodense surrogate markers around the three axes Y, Z, and X were on average 0.09, −0.52, and −0.01° with standard deviations of 2.01, 2.30, and 3.95°, respectively. The systematic uncertainty per patient for prostate rotation was estimated with 2.30, 1.56, and 4.13° and the mean random components with 1.81, 2.02, and 3.09°. The largest rotational errors occurred around the X-axis (pitch), but without preferring a certain orientation. Although the error around Z (roll) can be compensated on average by a transformation with 4 coordinates, a significant error around X remains and advocates the full correction with 6 coordinates. Rotational errors as assessed via daily stereoscopic online imaging are significant and dominate around X. Rotation possibly degrades the dosimetric coverage of the target volume and may require suitable strategies for correction. [Copyright &y& Elsevier]
Details
- Language :
- English
- ISSN :
- 09583947
- Volume :
- 37
- Issue :
- 4
- Database :
- Academic Search Index
- Journal :
- Medical Dosimetry
- Publication Type :
- Academic Journal
- Accession number :
- 82904227
- Full Text :
- https://doi.org/10.1016/j.meddos.2012.02.006