Back to Search Start Over

High rate read-out of LaBr(Ce) scintillator with a fast digitizer

Authors :
Stevanato, L.
Cester, D.
Nebbia, G.
Viesti, G.
Neri, F.
Petrucci, S.
Selmi, S.
Tintori, C
Source :
Nuclear Instruments & Methods in Physics Research Section A. Jun2012, Vol. 678, p83-87. 5p.
Publication Year :
2012

Abstract

Abstract: The energy resolution of a LaBr(Ce) detector has been studied as a function of the count rate up to 340kHz by using a 12 bit 250MS/s V1720 digitizer. The time resolution achieved by processing off line the digitized signals has been also determined. It appears that the energy resolution obtained with the digitizer is better than that achievable using standard NIM electronics. The time resolution yielded by the digitizer with a software CFTD is about δt=0.8ns (FWHM), slightly worse with respect to δt=0.65ns (FWHM) obtained from standard NIM. However, this time resolution lies well within the requirements for applications in Non-Destructive Analysis of large objects with tagged neutron beams. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
01689002
Volume :
678
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
75010362
Full Text :
https://doi.org/10.1016/j.nima.2012.03.016