Back to Search
Start Over
Upgraded photon calorimeter with integrating readout for the Hall A Compton polarimeter at Jefferson Lab
- Source :
-
Nuclear Instruments & Methods in Physics Research Section A . Jun2012, Vol. 676, p96-105. 10p. - Publication Year :
- 2012
-
Abstract
- Abstract: The photon arm of the Compton polarimeter in Hall A of Jefferson Lab has been upgraded to allow for electron beam polarization measurements with better than 1% accuracy. The data acquisition (DAQ) system now includes an integrating mode, which eliminates several systematic uncertainties inherent in the original counting-DAQ setup. The photon calorimeter has been replaced with a Ce-doped Gd2SiO5 crystal, which has a bright output and fast response, and works well for measurements using the new integrating method at electron beam energies from 1 to 6GeV. [Copyright &y& Elsevier]
Details
- Language :
- English
- ISSN :
- 01689002
- Volume :
- 676
- Database :
- Academic Search Index
- Journal :
- Nuclear Instruments & Methods in Physics Research Section A
- Publication Type :
- Academic Journal
- Accession number :
- 73987305
- Full Text :
- https://doi.org/10.1016/j.nima.2012.02.041