Back to Search Start Over

Upgraded photon calorimeter with integrating readout for the Hall A Compton polarimeter at Jefferson Lab

Authors :
Friend, M.
Parno, D.
Benmokhtar, F.
Camsonne, A.
Dalton, M.M.
Franklin, G.B.
Mamyan, V.
Michaels, R.
Nanda, S.
Nelyubin, V.
Paschke, K.
Quinn, B.
Rakhman, A.
Souder, P.
Tobias, A.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Jun2012, Vol. 676, p96-105. 10p.
Publication Year :
2012

Abstract

Abstract: The photon arm of the Compton polarimeter in Hall A of Jefferson Lab has been upgraded to allow for electron beam polarization measurements with better than 1% accuracy. The data acquisition (DAQ) system now includes an integrating mode, which eliminates several systematic uncertainties inherent in the original counting-DAQ setup. The photon calorimeter has been replaced with a Ce-doped Gd2SiO5 crystal, which has a bright output and fast response, and works well for measurements using the new integrating method at electron beam energies from 1 to 6GeV. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
01689002
Volume :
676
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
73987305
Full Text :
https://doi.org/10.1016/j.nima.2012.02.041