Back to Search Start Over

Characterization of the H3D ASIC Readout System and 6.0 cm^3 3-D Position Sensitive CdZnTe Detectors.

Authors :
Zhang, Feng
Herman, Cedric
He, Zhong
De Geronimo, Gianluigi
Vernon, Emerson
Fried, Jack
Source :
IEEE Transactions on Nuclear Science. 1/1/2012 Part 2 Part 2, Vol. 59 Issue 1, p236-242. 7p.
Publication Year :
2012

Abstract

Two 20 mm\,\times\,20 mm\,\times\,15 mm pixelated CZT detectors made by eV-Products were characterized using the new H3D Application Specific Integrated Circuits (ASIC) readout system developed by the Instrumentation Division at Brookhaven National Laboratory. The ASIC is capable of reading out energy and timing signals from 121 anode pixels and the planar cathode electrode of one CZT detector simultaneously. The system has a measured electronic noise of \sim \ 2.2 keV FWHM with a dynamic range from 20 keV to 3.0 MeV. The two detectors achieved energy resolution of 0.48% FWHM and 0.60% FWHM, respectively, at 662 keV for single-pixel events from the entire 6.0 cm^3 detection volume at room temperature with an un-collimated ^137Cs source. The average (\mu \tau)e of both detectors were measured to be >10^-2 cm^2/V. The detection efficiency of the two detectors was evaluated at several different energies up to 1.3 MeV by comparing with simulated data. It was found that the total counts agree well between the measured data and the simulated data over the studied energy range. However, the measured photopeak counts were 10–15% lower than simulated photopeak counts at high gamma-ray energies. The analysis shows that the loss of photopeak efficiency is likely due to the charge loss from peripheral pixels to the boundary of detectors. [ABSTRACT FROM PUBLISHER]

Details

Language :
English
ISSN :
00189499
Volume :
59
Issue :
1
Database :
Academic Search Index
Journal :
IEEE Transactions on Nuclear Science
Publication Type :
Academic Journal
Accession number :
73613944
Full Text :
https://doi.org/10.1109/TNS.2011.2175948