Back to Search Start Over

A 2.45 GHz electron cyclotron resonance proton ion source and a dual-lens low energy beam transport.

Authors :
Zhang, W. H.
Ma, H. Y.
Yang, Y.
Wu, Q.
Zhang, X. Z.
Wang, H.
Ma, B. H.
Feng, Y. C.
Fang, X.
Guo, J. W.
Cao, Y.
Li, X. X.
Zhu, Y. H.
Li, J. Y.
Sha, S.
Lu, W.
Lin, S. H.
Guo, X. H.
Zhao, H. Y.
Sun, L. T.
Source :
Review of Scientific Instruments. Feb2012, Vol. 83 Issue 2, p02A329. 3p. 4 Diagrams, 4 Graphs.
Publication Year :
2012

Abstract

The structure and preliminary commissioning results of a new 2.45 GHz ECR proton ion source and a dual-lens low energy beam transport (LEBT) system are presented in this paper. The main magnetic field of the ion source is provided by a set of permanent magnets with two small electro-solenoid magnets at the injection and the extraction to fine tune the magnetic field for better microwave coupling. A 50 keV pulsed proton beam extracted by a three-electrode mechanism passes through the LEBT system of length of 1183 mm. This LEBT consists of a diagnosis chamber, two Glaser lenses, two steering magnets, and a final beam defining cone. A set of inner permanent magnetic rings is embedded in each of the two Glaser lenses to produce a flatter axial-field to reduce the lens aberrations. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00346748
Volume :
83
Issue :
2
Database :
Academic Search Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
72107595
Full Text :
https://doi.org/10.1063/1.3669802