“MR conditional” respiratory ventilator system incident in a 3-T MRI environment

Authors :
Greenberg, Kathryn L.Z.
Weinreb, Jeffrey
Shellock, Frank G.
Source :
Magnetic Resonance Imaging (0730725X). Oct2011, Vol. 29 Issue 8, p1150-1154. 5p.
Publication Year :
2011

Abstract

Abstract: The misunderstanding of the labeling for an “MR conditional” respiratory ventilator system resulted in the placement of this device, which had substantial ferromagnetic components, too close to a 3-T magnetic resonance (MR) system, causing a projectile incident. Magnetic resonance imaging health care professionals should be aware of the potentially dangerous consequences when such medical devices that have Food and Drug Administration-approved MR conditional labeling are brought into the MR system room. Recommendations to prevent future incidents are provided herein. [Copyright &y& Elsevier]