Back to Search Start Over

LiF crystals as high spatial resolution neutron imaging detectors

Authors :
Matsubayashi, M.
Faenov, A.
Pikuz, T.
Fukuda, Y.
Kato, Y.
Yasuda, R.
Iikura, H.
Nojima, T.
Sakai, T.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Sep2011, Vol. 651 Issue 1, p90-94. 5p.
Publication Year :
2011

Abstract

Abstract: Neutron imaging by color center formation in LiF crystals was applied to a sensitivity indicator (SI) as a standard samples for neutron radiography. The SI was exposed to a 5mm pinhole-collimated thermal neutron beam with an LiF crystal and a neutron imaging plate (NIP) for 120min in the JRR-3M thermal neutron radiography facility. The image in the LiF crystal was read out using a laser confocal microscope. All gaps were clearly observed in images for both the LiF crystal and the NIP. The experimental results showed that LiF crystals have excellent characteristics as neutron imaging detectors in areas such as high spatial resolution. [Copyright &y& Elsevier]