Back to Search Start Over

Position Sensitive Detectors Mounted with Scintillators and Silicon Photomultipliers.

Authors :
Carvalhaes, Roberto P. M.
Bonifácio, Daniel A. B.
Moralles, Maurício
Source :
AIP Conference Proceedings. 8/10/2011, Vol. 1351 Issue 1, p202-206. 5p.
Publication Year :
2011

Abstract

This work presents the first results obtained in the 'Assembly and characterization of position sensitive detectors composed of scintillators coupled to silicon photomultipliers' project. The development of new x and γ radiation detectors have found several technological applications, especially in medical physics, where γ detectors that can be used in high intensity magnetic field are of particular importance. The experimental setup consisted of coupling of two silicon photomultipliers (SiPM) to the small sides of a 3×3×100 mm3 scintillator and the coupling of one SiPM to one of the small sides of a 3×3×10 mm3 scintillator. We found that the detectors used in this study presented an energy resolution that is in agreement with those observed in scintillators of the same family coupled to conventional photomultipliers. Besides that, there is a strong correlation between the difference of the light intensity in both SiPMs of the long detector and the position of the γ source. The results confirm the great potential of application of such detectors. [ABSTRACT FROM AUTHOR]