Back to Search Start Over

Measuring low levels of protein aggregation by sedimentation velocity

Authors :
Gabrielson, John P.
Arthur, Kelly K.
Source :
Methods. May2011, Vol. 54 Issue 1, p83-91. 9p.
Publication Year :
2011

Abstract

Abstract: The required performance of an analytical method depends on the purpose for which it will be used. As a methodology matures, it may find new application, and the performance demands placed on the method can increase. Sedimentation velocity analytical ultracentrifugation (SV-AUC) has a long and distinguished history with important contributions to molecular biology. Now the technique is transitioning into industrial settings, and among them, SV-AUC is now used to quantify the amount of protein aggregation in biopharmaceutical protein products, often at levels less than 1% of the total protein mass. In this paper, we review recent advances to SV methodology which have been shown to improve quantitation of protein aggregation. Then we discuss the performance of the SV method in its current state, with emphasis on the precision and quantitation limit of the method, in the context of existing industrial guidance on analytical method performance targets for quantitative methods. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
10462023
Volume :
54
Issue :
1
Database :
Academic Search Index
Journal :
Methods
Publication Type :
Academic Journal
Accession number :
60517389
Full Text :
https://doi.org/10.1016/j.ymeth.2010.12.030