Back to Search Start Over

X-ray micro-tomography-based non-destructive evaluation of microwelds in an HDR source assembly used for brachytherapy.

Authors :
Sarkar, P. S.
Sinha, A.
Mishra, M. K.
Badodkar, D. N.
Singh, M.
Sastry, K. V. S.
Kohli, A. K.
Source :
Insight: Non-Destructive Testing & Condition Monitoring. Jan2011, Vol. 53 Issue 1, p25-28. 4p. 14 Black and White Photographs.
Publication Year :
2011

Abstract

The article presents a study on machinery testing of high dose rate (HDR) source assembly for brachytherapy through a non-destructive process, three-dimensional (3D) X-ray micro-tomography. It states that the study aims to standardise the welding parameters for the development of the HRD source assembly. It also proves the effectiveness of 3D X-ray micro-tomography for detecting micro-welds defects in HDR source assemblies.

Details

Language :
English
ISSN :
13542575
Volume :
53
Issue :
1
Database :
Academic Search Index
Journal :
Insight: Non-Destructive Testing & Condition Monitoring
Publication Type :
Academic Journal
Accession number :
57434474
Full Text :
https://doi.org/10.1784/insi.2011.53.1.25