Back to Search
Start Over
Disposable biosensor based on immobilisation of glutamate oxidase on Pt nanoparticles modified Au nanowire array electrode
- Source :
-
Biosensors & Bioelectronics . Dec2010, Vol. 26 Issue 4, p1420-1424. 5p. - Publication Year :
- 2010
-
Abstract
- Abstract: Novel electrochemical platform based on Pt nanoparticle modified ordered three-dimensional gold nanowire arrays (PtNP/NAEs) for the amperometric sensing of H2O2 and glutamate is developed. Pt nanoparticle (PtNP) is fabricated by electrodeposition onto the 3D nanowires and characterised using scanning electron microscopy (SEM) and cyclic voltammetry. The deposited nanoparticles have an average size of 20nm. The PtNP/NAE shows a linear response of up to 20mM for H2O2 detection with a sensitivity of 194.60μAmM−1 cm−2 at 20°C. It can detect 1μM (S/N=3) of H2O2 at normal condition without using any enzyme or mediator. Analytical performance of this electrode is tested by immobilising glutamate oxidase (GlutOx) through cross-linking in the matrix of bovine serum albumin (BSA), Nafion and glutaraldehyde. At physiological pH, the biosensor showed the sensitivity of 10.76μAmM−1 cm−2, with a linear range of up to 0.8mM. [ABSTRACT FROM AUTHOR]
- Subjects :
- *BIOSENSORS
*GLUTAMIC acid
*PLATINUM compounds
*NANOPARTICLES
*NANOWIRES
*ELECTRODES
Subjects
Details
- Language :
- English
- ISSN :
- 09565663
- Volume :
- 26
- Issue :
- 4
- Database :
- Academic Search Index
- Journal :
- Biosensors & Bioelectronics
- Publication Type :
- Academic Journal
- Accession number :
- 55499146
- Full Text :
- https://doi.org/10.1016/j.bios.2010.07.071