Back to Search Start Over

Characterization of charge carrier collection in a CdZnTe Frisch collar detector with a highly collimated source

Authors :
Kargar, Alireza
Harrison, Mark J.
Brooks, Adam C.
McGregor, Douglas S.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Aug2010, Vol. 620 Issue 2/3, p270-278. 9p.
Publication Year :
2010

Abstract

Abstract: A 4.7 ×4.7×9.5mm3 CdZnTe Frisch collar device was characterized through probing the device with a highly collimated 662keV gamma ray source. In a systematic series of experiments, the detector was probed along the length and width with a gamma ray source using a 43.0mm long Pb-collimator with a 0.6mm circular hole. The detector was probed along the central line under different operating voltages of 1200, 1000, 800, 600 and 400V. The experimental results correlated well to charge collection calculations for a modeled device with the same size and operating conditions. It was proved that, unlike the planar configuration, the charge collection efficiency profile along the length of Frisch collar device is considerably improved. The CdZnTe raw materials for this study were acquired from Redlen Technologies, and the Frisch collar device was fabricated and characterized at S.M.A.R.T. Laboratory at Kansas State University. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01689002
Volume :
620
Issue :
2/3
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
52347163
Full Text :