Back to Search Start Over

NOVEL DESIGN FOR CENTRIFUGAL COUNTER-CURRENT CHROMATOGRAPHY: III. SAW TOOTH COLUMN.

Authors :
Yang, Yi
Aisa, HajiAkber
Ito, Yoichiro
Source :
Journal of Liquid Chromatography & Related Technologies. 2010, Vol. 33 Issue 6, p846-858. 13p. 2 Black and White Photographs, 3 Charts, 4 Graphs.
Publication Year :
2010

Abstract

The toroidal coil using an equilateral triangular core and zigzag pattern column have improved both retention of the stationary phase and peak resolution of the conventional toroidal coil in centrifugal counter-current chromatography. To further improve the retention of stationary phase and peak resolution, a novel saw tooth column was designed and the performance of the system was evaluated at various flow rates. The results indicated that both retention of the stationary phase and peak resolution were improved as the flow rate was decreased and at a flow rate of 0.005 mL/min the resolution is remarkably increased. Modification of the tubing called flat-twisted tubing further improved the peak resolution without increasing the column pressure. With decreased column length at a capacity of about 0.2 mL, resolution of the saw tooth column was 1.02. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
10826076
Volume :
33
Issue :
6
Database :
Academic Search Index
Journal :
Journal of Liquid Chromatography & Related Technologies
Publication Type :
Academic Journal
Accession number :
48795576
Full Text :
https://doi.org/10.1080/10826071003684547