Photon W-value of dry air determined using a cryogenic radiometer combined with a multi-electrode ion chamber for soft X-rays
- Source :
-
Radiation Physics & Chemistry . Apr2010, Vol. 79 Issue 4, p397-404. 8p. - Publication Year :
- 2010
-
Abstract
- Abstract: The photon W-value of dry air in the energy range of 2.2–3.9keV has been measured with relative expanded uncertainty of 0.84% (coverage factor ). Monochromatized synchrotron radiation was used for both measurements of the radiant power by means of a cryogenic radiometer and of photoion currents for ion-pair production by means of a multi-electrode ion chamber technique. The evaluated dependence of the photon W-value for dry air is consistent with the previously reported data for electrons and seems to show a small change at the Ar 1s ionization threshold. These findings were well reproduced with a model calculation using atomic photoionization data. Mass attenuation coefficients for air have been observed in the present soft X-ray region with a small energy step, which are close to data previously reported. [Copyright &y& Elsevier]