Model za usporedivost rezultata dobivenih s dva različita biokemijska analizatora u laboratoriju akreditiranom prema normi ISO 15189.

Authors :
Miler, Marijana
Šimundić, Ana-Maria
Štefanović, Mario
Ferenec-Ružić, Dragica
Kvaternik, Marina
Topić, Elizabeta
Vrkić, Nada
Source :
Biochemia Medica. Dec2009, Vol. 19 Issue 3, p287-293. 7p. 1 Diagram, 2 Charts.
Publication Year :
2009

Abstract

Introduction: Laboratories accredited according to ISO 15189 standard which have two or more different analyzers or procedures for examinations, should define mechanism for comparison of results. The objective of this study was to present the model for comparison of emergency tests on two different types of automated general biochemistry analyzers. Materials and methods: General biochemistry emergency tests were measured in serum and urine on two automated analyzers Olympus AU2700 and Olympus AU640 (Olympus, Hamburg, Germany) during 60 day period. Olympus AU2700 was used as the reference analyzer and Olympus AU640 as a backup system. Control samples on two levels were used for result comparison. Results: Average 60-day period bias was within acceptance criteria for all analyzes included in this model of comparability assessment. Only low control concentrations for some tests were exceeding allowable biases in less than 2% cases and high level controls were never exceeding allowable biases. Conclusions: This model of results comparison ensures continuous analytical quality of laboratory reports and improves overall efficiency of laboratory services. [ABSTRACT FROM AUTHOR]

Details

Language :
Croatian
ISSN :
13300962
Volume :
19
Issue :
3
Database :
Academic Search Index
Journal :
Biochemia Medica
Publication Type :
Academic Journal
Accession number :
47186416