Back to Search Start Over

A setup for thermodesorption measurements.

Authors :
Rusinov, A. A.
Gasparyan, Yu. M.
Perelygin, S. F.
Pisarev, A. A.
Stepanov, S. O.
Trifonov, N. N.
Source :
Instruments & Experimental Techniques. Nov2009, Vol. 52 Issue 6, p871-876. 6p. 2 Diagrams, 3 Graphs.
Publication Year :
2009

Abstract

An ultrahigh-vacuum setup for studying the interaction of ions with a solid by the thermodesorption spectroscopy method has been developed. The residual pressure achieved in the setup after heating is <2 × 10−7 Pa (the main component of the residual gas is H2). A sample suspended by two tungsten wires is inserted into the main vacuum chamber through a lock chamber. The sample is heated by thermal radiation from the heater from room temperature to 1600–1700 K. The main features of the setup, the experimental technique, and the calibration of desorption flows are discussed. Examples of results obtained on the setup are presented. [ABSTRACT FROM AUTHOR]