Estudo comparativo da qualidade de imagem dos modos de aquisição da PET: validação de um protocolo para reduzir a dose de radiação.

Authors :
Nogueira, Solange Amorim
Lederman, Henrique Manoel
Wagner, Jairo
Yamaga, Lilian Yuri
da Cunha, Marcelo Livorsi
de Gusmão Funari, Marcelo Buarque
Source :
Radiologia Brasileira. mar/abr2009, Vol. 42 Issue 2, p103-108. 6p. 2 Diagrams, 4 Charts, 2 Graphs.
Publication Year :
2009

Abstract

OBJECTIVE: The present study is aimed at validating a 3D acquisition protocol for positron emission tomography as a replacement for the 2D mode, to reduce the radiation dose delivered to patients, without any loss in the quality of images. MATERIALS AND METHODS: The study comprised 27 simulations in a Discovery ST equipment with four-minute 2D acquisitions, and two-minute and four-minute 3D acquisitions, utilizing a chest phantom according to the National Electrical Manufacturers Association protocol. Six spheres with different diameters were inserted into this phantom as a means for determining the images quality. The images were blindly reviewed by three experienced nuclear physicians who did not know the acquisition modes. Each observer attributed a score 1 when one of the spheres was not identified, or 2 for visible spheres. RESULTS: The qualitative analysis based on generalized kappa coefficient demonstrated that the frequency of visible spheres was higher with four-minute 3D acquisitions (85%), with an also higher interobserver agreement (88.9%), generalized kappa = 0.725 [0.507;0.942]. CONCLUSION: The protocol with four-minute 3D acquisition with lower 18F-FDG activity can be utilized for patients with a biotype similar to the phantom, without any loss in the imaging quality. [ABSTRACT FROM AUTHOR]

Details

Language :
Portuguese
ISSN :
01003984
Volume :
42
Issue :
2
Database :
Academic Search Index
Journal :
Radiologia Brasileira
Publication Type :
Academic Journal
Accession number :
44286787
Full Text :
https://doi.org/10.1590/S0100-39842009000200008