Back to Search
Start Over
X-ray detection with 3D Medipix2 devices
- Source :
-
Nuclear Instruments & Methods in Physics Research Section A . Aug2009, Vol. 607 Issue 1, p89-91. 3p. - Publication Year :
- 2009
-
Abstract
- Abstract: A set of double-sided 3D silicon detectors have been fabricated, bump bonded to single-photon-counting Medipix2 readout chips, and tested using X-ray sources. When tested with monochromatic X-rays, the 3D detectors show reduced charge sharing compared with standard planar pixel detectors [Copyright &y& Elsevier]
Details
- Language :
- English
- ISSN :
- 01689002
- Volume :
- 607
- Issue :
- 1
- Database :
- Academic Search Index
- Journal :
- Nuclear Instruments & Methods in Physics Research Section A
- Publication Type :
- Academic Journal
- Accession number :
- 43353350
- Full Text :
- https://doi.org/10.1016/j.nima.2009.03.135