Back to Search Start Over

Simple dose rate measurements for a very high synchrotron X-ray flux.

Authors :
Chi-Jen Liu
Chang-Hai Wang
Cheng-Liang Wang
Hwu, Y.
Chien-Yi Lin
Margaritondo, G.
Source :
Journal of Synchrotron Radiation. May2009, Vol. 16 Issue 3, p395-397. 3p.
Publication Year :
2009

Abstract

Dose measurements based on methylene blue (MB) bleaching, widely used for ultraviolet light, can also be applied to X-rays including very high flux levels. This method has been tested by using both MB bleaching and Fricke dosimetry for a conventional monochromatic X-ray source and then for 'white-beam' synchrotron radiation. The results show that MB bleaching dosimetry can easily measure X-ray doses up to at least 10[sup5] Gy s[sup-1], as long as the MB concentration is sufficiently high. This condition can be verified from the deviations from linearity of the bleaching versus exposure time. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09090495
Volume :
16
Issue :
3
Database :
Academic Search Index
Journal :
Journal of Synchrotron Radiation
Publication Type :
Academic Journal
Accession number :
40643166
Full Text :
https://doi.org/10.1107/S0909049509007225