Abstract
- Dose measurements based on methylene blue (MB) bleaching, widely used for ultraviolet light, can also be applied to X-rays including very high flux levels. This method has been tested by using both MB bleaching and Fricke dosimetry for a conventional monochromatic X-ray source and then for 'white-beam' synchrotron radiation. The results show that MB bleaching dosimetry can easily measure X-ray doses up to at least 10[sup5] Gy s[sup-1], as long as the MB concentration is sufficiently high. This condition can be verified from the deviations from linearity of the bleaching versus exposure time. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 09090495
- Volume :
- 16
- Issue :
- 3
- Database :
- Academic Search Index
- Journal :
- Journal of Synchrotron Radiation
- Publication Type :
- Academic Journal
- Accession number :
- 40643166
- Full Text :
- https://doi.org/10.1107/S0909049509007225