Humidification Performance of 48 Passive Airway Humidifiers: Comparison With Manufacturer Data.

Authors :
Lellouche, François
Taillé, Solenne
Lefrançois, Frédéric
Deye, Nicolas
Maggiore, Salvatore Maurizio
Jouvet, Philippe
Ricard, Jean-Damien
Fumagalli, Bruno
Brochard, Laurent
Source :
CHEST. Feb2009, Vol. 135 Issue 2, p276-286. 11p.
Publication Year :
2009

Abstract

The article presents a study to evaluate the humidification performance of a large number of adult heat and moisture exchangers (HMEs). It assesses 48 devices using a bench test apparatus that simulated real-life physiologic ventilation conditions, and performs three hygrometric measurements for each device. The research suggests that several HMEs performed poorly and should not be used as HMEs.