Back to Search Start Over

Development of a high performance liquid chromatography method for quantification of PAC-1 in rat plasma

Authors :
Fang, Li-Na
Chen, Xiao-Hui
Song, Zhen
Wang, Gang
Zhao, Xu
Ren, Lei
Gong, Ping
Bi, Kai-Shun
Source :
Journal of Pharmaceutical & Biomedical Analysis. Feb2009, Vol. 49 Issue 2, p447-450. 4p.
Publication Year :
2009

Abstract

Abstract: A sensitive and specific high performance liquid chromatography method with UV detection was developed and validated for the determination of PAC-1 in rat plasma. After extraction with ethyl acetate, the chromatographic separation was carried out on a Diamonsil C18 column (150mm×4.6mm i.d., 5μm particle size, Zhonghuida) protected by a ODS guard column (10mm×4.6mm i.d., 5μm particle size), using acetonitrile–methanol–phosphate buffer (pH 3.0, 30mM) (31:3:66, v/v/v) as mobile phase at a flow rate of 1.0mL/min, and wavelength of the UV detector was set at 281nm. No interference from any endogenous substances was observed during the elution of PAC-1 and internal standard (IS, indapamide). The calibration curve was linear over the range of 0.05–20μg/mL (r >0.99). The lower limit of quantification was evaluated to be 50ng/mL. The method was successfully applied to the pharmacokinetic study of PAC-1 after intravenous and oral administration in rats, respectively. [Copyright &y& Elsevier]