Back to Search
Start Over
Total characterization of neutron detectors with a 252Cf source and a new light output determination
- Source :
-
Nuclear Instruments & Methods in Physics Research Section A . Feb2009, Vol. 599 Issue 2/3, p226-233. 8p. - Publication Year :
- 2009
-
Abstract
- Abstract: A full characterization of neutron detector properties was achieved by means of a 252Cf source with timing of fission events. The input neutron energies were separated with the time-of-flight technique. The pulse height, the pulse shape and the time-of-flight spectra were measured in coincidence mode as three-dimensional distributions. On the basis of these data, the following characteristics were determined for each detector: response function, light output versus proton (neutron) energy, energy resolution function for neutron energies from 1 to 7MeV, time shift correction, and finally the detector efficiency from threshold up to 12MeV. The experimental results were compared with Monte-Carlo simulations. Two new analytical dependencies are suggested for the light output calculation for electrons and protons. [Copyright &y& Elsevier]
Details
- Language :
- English
- ISSN :
- 01689002
- Volume :
- 599
- Issue :
- 2/3
- Database :
- Academic Search Index
- Journal :
- Nuclear Instruments & Methods in Physics Research Section A
- Publication Type :
- Academic Journal
- Accession number :
- 36189475
- Full Text :
- https://doi.org/10.1016/j.nima.2008.10.032