Abstract
- Abstract: In this study, the reliability of thermal desorbing technique was investigated using the gaseous standards of five reduced sulfur compounds (RSCs: hydrogen sulfide, methane thiol, dimethyl sulfide, carbon disulfide, and dimethyl disulfide). A series of calibration experiments for RSCs were performed using gas chromatography (GC) with pulsed flame photometric detector (PFPD) that is interfaced with a thermal desorber (TD) unit. These calibration data were evaluated by means of two contrasting concepts: fixed standard concentration method (FSC: variable volumetric injection of standard gases prepared at a given concentration) and fixed standard volume method (FSV: injection of multiple standards with varying concentrations at a given volume). When the results of both methods were compared, RSCs generally showed enhanced sensitivity with increasing concentration (FSC) and sample loading volume (FSV). This study highlights that TD-based calibration properties are practically undistinguishable between different sample transfer approaches (e.g., FSV and FSC). As a result, the calibration properties of RSCs derived by thermal desorption technique are greatly distinguished from those of direct injection into GC. [Copyright &y& Elsevier]
Details
- Language :
- English
- ISSN :
- 0026265X
- Volume :
- 91
- Issue :
- 1
- Database :
- Academic Search Index
- Journal :
- Microchemical Journal
- Publication Type :
- Academic Journal
- Accession number :
- 35562990
- Full Text :
- https://doi.org/10.1016/j.microc.2008.07.004