Comparison of four extraction methods for the analysis of 24 pesticides in soil samples with gas chromatography–mass spectrometry and liquid chromatography–ion trap–mass spectrometry

Authors :
Lesueur, C.
Gartner, M.
Mentler, A.
Fuerhacker, M.
Source :
Talanta. Mar2008, Vol. 75 Issue 1, p284-293. 10p.
Publication Year :
2008

Abstract

Abstract: With a view to analyse multiresidues of pesticides in soil samples, a new ultrasonic solvent extraction (USE) was compared to the European Norm DIN 12393 for foodstuff (extraction with acetone, partitioning with ethylacetate/cyclohexane and clean-up with gel permeation chromatography (GPC)), the QuECheRS method and a pressurised liquid extraction (PLE) method. Pesticides were analysed with both GC-MS and HPLC-MS/MS. The reference materials were the EUROSOIL 7 and its subsoil SO26 as well as a sea sand. All the substances were observed to be linear in the range of 4–800ngg−1 for the European Norm DIN 12393, 7–1400ngg−1 for the USE method and 20–4000ngg−1 for the QueCheRS and the PLE methods. Limits of detection (LOD) and limits of quantification (LOQ) were with HPLC between 0.006 and 0.23ngg−1 and between 0.022 and 0.77ngg−1, respectively, with the exception of diuron (LOD up to 11.8ngg−1; LOQ up to 39.2ngg−1). With GC they ranged from 3.0 to 87.5ngg−1 and from 10 to 292ngg−1, respectively. All substances could be recovered with USE as well as with the QuECheRS method; the European Norm DIN 12393 could not recover carbendazim and metamitron; the PLE carbendazim, metamitron and monolinuron. For the remaining substances, recoveries at a 500ngg−1 fortification level ranged from 10.9 to 96.3% with the USE. In comparison, the QuECheRS method was the most efficient extraction method with recoveries from 27.3 to 120.9%. It was followed by the European Norm DIN 12393 with recoveries between 6.8 and 108.1% and the PLE with recoveries from 12.2 to 153.2%. Recoveries were higher from the EUROSOIL 7 than from the SO 26. The repeatability expressed in term of standard deviation was below 20% for all substances and all materials. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
00399140
Volume :
75
Issue :
1
Database :
Academic Search Index
Journal :
Talanta
Publication Type :
Academic Journal
Accession number :
30615488
Full Text :
https://doi.org/10.1016/j.talanta.2007.11.031