Back to Search Start Over

Determination of scopoletin in rat plasma by high performance liquid chromatographic method with UV detection and its application to a pharmacokinetic study

Authors :
Xia, Yufeng
Dai, Yue
Wang, Qiang
Liang, Huizheng
Source :
Journal of Chromatography B: Analytical Technologies in the Biomedical & Life Sciences. Oct2007, Vol. 857 Issue 2, p332-336. 5p.
Publication Year :
2007

Abstract

Abstract: A rapid and simple high-performance liquid chromatographic (HPLC) method has been developed and validated for determination of scopoletin in rat plasma using psoralen as internal standard. Chromatographic separation was achieved on a C18 column using methanol and distilled water (49:51, v/v) containing 0.05% (v/v) phosphoric acid as mobile phase. The UV detector was set at 345nm. The calibration curve was linear over the range of 0.165–9.90μg/ml with a correlation coefficient of 0.9994. The recovery for plasma samples of 0.165, 1.32 and 6.60μg/ml was 93.2%, 95.9% and 95.5%, respectively. The RSD of intra- and inter-day assay variations was less than 6.7%. This HPLC assay is a precise and reliable method for the analysis of scopoletin in pharmacokinetic studies. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
15700232
Volume :
857
Issue :
2
Database :
Academic Search Index
Journal :
Journal of Chromatography B: Analytical Technologies in the Biomedical & Life Sciences
Publication Type :
Academic Journal
Accession number :
26839844
Full Text :
https://doi.org/10.1016/j.jchromb.2007.07.023