Back to Search
Start Over
Low-level quantitation of oxycodone and its oxidative metabolites, noroxycodone, and oxymorphone, in rat plasma by high-performance liquid chromatography–electrospray ionization–tandem mass spectrometry
- Source :
-
Journal of Chromatography B: Analytical Technologies in the Biomedical & Life Sciences . Apr2007, Vol. 848 Issue 2, p264-270. 7p. - Publication Year :
- 2007
-
Abstract
- Abstract: A method was developed for quantification of oxycodone, noroxycodone, and oxymorphone in small volumes (50μl) of rat plasma by high-performance liquid chromatography–electrospray ionization–tandem mass spectrometry using turbo ion-spray. Deuterated (d3) opioid analogues acted as internal standards. Sample preparation involved protein precipitation with acetonitrile, centrifugal evaporation, and reconstitution in mobile phase; analyte separation was performed on a C18 (5μm, 2.1mm×50mm) column using a linear gradient program. Lower limits of quantitation (ng/ml) and their between-day accuracy and precision were—oxycodone, 0.9 (−0.2 and 7.8%); noroxycodone, 1.0 (0.6 and 6.2%); oxymorphone 1.0 (−1.8 and 9.5%). [Copyright &y& Elsevier]
Details
- Language :
- English
- ISSN :
- 15700232
- Volume :
- 848
- Issue :
- 2
- Database :
- Academic Search Index
- Journal :
- Journal of Chromatography B: Analytical Technologies in the Biomedical & Life Sciences
- Publication Type :
- Academic Journal
- Accession number :
- 24461234
- Full Text :
- https://doi.org/10.1016/j.jchromb.2006.10.039