Back to Search
Start Over
A rapid, simple measurement of human albumin in whole blood using a fluorescence immunoassay (I)
- Source :
-
Clinica Chimica Acta . Jan2004, Vol. 339 Issue 1/2, p147-156. 10p. - Publication Year :
- 2004
-
Abstract
- Background: Human serum albumin (HSA) is the most abundant plasma protein and plays key a role in metabolism. The variation in albumin concentration provides valuable information related to metabolic diseases and diagnostic application. Methods: We constructed two assay systems to quantify the albumin concentration. The immunoassay used a fluorescence (FL) dye to detect albumin in samples and employed the conventional chromatography as a separation system. The assay system consists of an anti-HSA-mAb or an HSA immobilized test strip in a disposable cartridge, a fluorescence-labeled detector buffer and a laser-fluorescence scanner. We mixed the sample with detector, loaded it onto a cartridge, incubated it for 10 min and measured the concentration of albumin in a laser-fluorescence scanner. We examined the comparability of assay with an automated BCG dye binding method using a Hitachi 747 biochemical analyzer. Results: The correlation of coefficient between AT/AC as converted from the relative fluorescence units (RFU) and albumin concentration displayed reasonable reliability in both the competition and the inhibition assay systems (r=0.998). Using the Bland–Altman difference plot analysis, we observed an acceptable agreement between two methods, the fluorescence immunochromatography assay (FL-ICA) and the automated BCG dye-binding method of a Hitachi biochemical analyzer, over the clinical relevant range of HSA concentrations. The coefficient of variation (CV) of within- and between-run variation in the immunoassay system was <8% and the recovery fell within 5% in each control sample. In addition to its reliable analytical performance, the assay with whole blood can be completed in 12 min using a one-step operation without any pretreatment. Conclusion: The developed immunoassay system using fluorescence dye and lateral-flow chromatography is a simple, fast and reliable method for quantifying the albumin concentration in whole blood. [Copyright &y& Elsevier]
- Subjects :
- *IMMUNOASSAY
*ALBUMINS
*FLUORESCENCE
*BLOOD plasma
Subjects
Details
- Language :
- English
- ISSN :
- 00098981
- Volume :
- 339
- Issue :
- 1/2
- Database :
- Academic Search Index
- Journal :
- Clinica Chimica Acta
- Publication Type :
- Academic Journal
- Accession number :
- 22238744
- Full Text :
- https://doi.org/10.1016/j.cccn.2003.10.002