Back to Search
Start Over
Optimised filtering for improved energy and position resolution in position-sensitive TES based X-ray detectors
- Source :
-
Nuclear Instruments & Methods in Physics Research Section A . Jan2006, Vol. 556 Issue 1, p237-245. 9p. - Publication Year :
- 2006
-
Abstract
- Abstract: We report on a new optimal filtering algorithm for improved energy and position resolution in position sensitive Transition Edge Sensor (TES) based X-ray detectors. This algorithm takes account of the noise correlation between the detector outputs to minimise the variance on the estimated photon energy and position. Using numerical methods we show that improved energy and position resolution can be obtained than from previously published methods. Our simulations also reveal the trade-offs resulting from changes in the thermal conductances and heat capacities of the detector elements. [Copyright &y& Elsevier]
- Subjects :
- *FINITE element method
*ALGORITHMS
*DETECTORS
*ENGINEERING instruments
Subjects
Details
- Language :
- English
- ISSN :
- 01689002
- Volume :
- 556
- Issue :
- 1
- Database :
- Academic Search Index
- Journal :
- Nuclear Instruments & Methods in Physics Research Section A
- Publication Type :
- Academic Journal
- Accession number :
- 19199094
- Full Text :
- https://doi.org/10.1016/j.nima.2005.10.021