Back to Search Start Over

Gain stability of Hamamatsu R5912-MOD photomultipliers at low temperature.

Authors :
Saia, C.
Brio, V.
Persiani, R.
Petta, C.M.A.
Raselli, G.L.
Rossella, M.
Vicenzi, M.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Dec2024, Vol. 1069, pN.PAG-N.PAG. 1p.
Publication Year :
2024

Abstract

The Photon Detection System (PDS) is a crucial component of the ICARUS detector in the Short Baseline Neutrino (SBN) Program at Fermi National Accelerator Laboratory (FNAL). It consists of 360 PhotoMultiplier Tubes (PMTs) 8" Hamamatsu R5912-MOD and, since June 2020, it has been operating at the liquid Argon cryogenic temperature. During these years, the PMTs have shown gain losses as an aging effect. Using a climatic chamber at the INFN Sezione di Catania, we confirmed a stable gain for a single PMT 8" Hamamatsu R5912-MOD when operated at room temperature, and a persistent gain loss at temperatures around -70 °C, although this temperature is rather higher than that of the liquid Argon. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01689002
Volume :
1069
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
180773530
Full Text :
https://doi.org/10.1016/j.nima.2024.169861