Abstract
- The Photon Detection System (PDS) is a crucial component of the ICARUS detector in the Short Baseline Neutrino (SBN) Program at Fermi National Accelerator Laboratory (FNAL). It consists of 360 PhotoMultiplier Tubes (PMTs) 8" Hamamatsu R5912-MOD and, since June 2020, it has been operating at the liquid Argon cryogenic temperature. During these years, the PMTs have shown gain losses as an aging effect. Using a climatic chamber at the INFN Sezione di Catania, we confirmed a stable gain for a single PMT 8" Hamamatsu R5912-MOD when operated at room temperature, and a persistent gain loss at temperatures around -70 °C, although this temperature is rather higher than that of the liquid Argon. [ABSTRACT FROM AUTHOR]