In-situ relative calibration of Bragg crystals with Monte Carlo line ratio analysis.

Authors :
Kraus, B. F.
Kriz, E.
McPoyle, S. P.
Atay, K.
Malko, S.
Hill, K. W.
Gao, Lan
Efthimion, P. C.
Hollinger, R. C.
Wang, Shoujun
King, J.
Zahedpour Anaraki, S.
Rocca, J. J.
Source :
Review of Scientific Instruments. Oct2024, Vol. 95 Issue 10, p1-6. 6p.
Publication Year :
2024

Abstract

X-ray line emission spectra can thoroughly characterize hot plasmas, especially when line shapes and ratios convey distinct aspects of plasma conditions. However, the high spectral resolution required for observing line shapes is often at odds with the large bandwidth required to observe many line ratios across a wide spectral range. One strategy to obtain high spectral resolution over a wide bandwidth is to use multiple crystals with calibrated reflectivity so that line intensities across different crystals can be compared. Here, we explore the use of a low-resolution, wide-bandwidth mica survey spectrometer to infer relative reflectivity of two high-resolution, narrow-bandwidth quartz crystals. A Monte Carlo error analysis determines comparable x-ray line ratios measured from both spectrometers, resulting in an in situ calibration factor and associated uncertainty for the relative reflectivity of the high-resolution crystals. [ABSTRACT FROM AUTHOR]