Back to Search Start Over

Quantifying the effects of neutron fluence on proton signal retention in CR-39.

Quantifying the effects of neutron fluence on proton signal retention in CR-39.

Authors :
Russell, L.
Johnson, T. M.
Lawrence, Y.
Reichelt, B.
Vanderloo, N.
Cufari, M.
Buschmann, B. I.
Dannhoff, S.
DeVault, A.
Doeg, E.
Evans, T.
Foo, B. C.
Frankel, R.
Kunimune, J. H.
Pearcy, J. A.
Vargas, J.
Gatu Johnson, M.
Frenje, J.
Source :
Review of Scientific Instruments. Oct2024, Vol. 95 Issue 10, p1-5. 5p.
Publication Year :
2024

Abstract

This paper reports on investigations on the impact of higher neutron fluences on the detection efficiency of protons with CR-39, a charged particle track detector. CR-39 is widely used as a diagnostic for inertial fusion applications and is an integral component of numerous particle diagnostics at the OMEGA laser facility and National Ignition Facility. As experiments continue to produce higher and higher yields, existing diagnostics are impacted by higher particle fluences than they were originally designed for. This paper presents data from experiments measuring proton signal on pieces of CR-39 with different levels of neutron fluence with two different etch times. The experiments show a decrease in signal recovery with increased neutron fluence, which is exacerbated at longer etch times. At 3 h etch time, data suggest a 17% ± 7% signal loss at 1.3 × 105 neutron-induced tracks per cm2 and a 67% ± 21% loss at 6 h etch time. Careful signal isolation techniques can recover most of the proton tracks even with moderate neutron fluence. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00346748
Volume :
95
Issue :
10
Database :
Academic Search Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
180631754
Full Text :
https://doi.org/10.1063/5.0219479