Back to Search Start Over

A new dual‐thickness semi‐transparent beamstop for small‐angle X‐ray scattering.

Authors :
Wu, Haijuan
Li, Zhihong
Source :
Journal of Synchrotron Radiation. Sep2024, Vol. 31 Issue 5, p1197-1208. 12p.
Publication Year :
2024

Abstract

An innovative dual‐thickness semi‐transparent beamstop designed to enhance the performance of small‐angle X‐ray scattering (SAXS) experiments is introduced. This design integrates two absorbers of differing thicknesses side by side into a single attenuator, known as a beamstop. Instead of completely stopping the direct beam, it attenuates it, allowing the SAXS detector to measure the transmitted beam through the sample. This approach achieves true synchronization in measuring both scattered and transmitted signals and effectively eliminates higher‐order harmonic contributions when determining the transmission light intensity through the sample. This facilitates and optimizes signal detection and background subtraction. This contribution details the theoretical basis and practical implementation of this solution at the SAXS station on the 1W2A beamline at the Beijing Synchrotron Radiation Facility. It also anticipates its application at other SAXS stations, including that at the forthcoming High Energy Photon Source, providing an effective solution for high‐precision SAXS experiments. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09090495
Volume :
31
Issue :
5
Database :
Academic Search Index
Journal :
Journal of Synchrotron Radiation
Publication Type :
Academic Journal
Accession number :
179411807
Full Text :
https://doi.org/10.1107/S1600577524007392