Back to Search Start Over

Clinical Utility of Recently Food and Drug Administration-Approved IntelliSep Test (Sepsis Biomarker) for Early Diagnosis of Sepsis: Comparison with Other Biomarkers.

Authors :
Sarani, Nima
Dasgupta, Amitava
Enders, Maria
Rowan, Lauren
Elsarraj, Hanan
Gralnek, Sarah
Shay, Madison
Lemar, Lucas R.
Simpson, Steven Q.
Cunningham, Mark T.
Zheng, X. Long
Source :
Journal of Clinical Medicine. Aug2024, Vol. 13 Issue 16, p4852. 9p.
Publication Year :
2024

Abstract

Context: IntelliSep by Cytovale has received United States (U.S.) Food and Drug Administration (FDA) approval as a sepsis biomarker test. However, the clinical utility of this new test is not assessed in emergency departments. Objective: We investigated the clinical utility of this test using 44 patients visiting the emergency department at The University of Kansas Medical Center by comparing it with the monocyte distribution width (MDW) and other biomarkers including the von Willebrand factor (vWF) and ADAMTS13. Design and Methods: IntelliSep assesses the cellular host response via deformability cytometry of biophysical leukocyte properties and produces a score (IntelliSep Index; ISI: from 0.1 (lowest risk) to 10 (highest risk). We measured the ISI in 44 patients (19 high probability and 25 low probability of sepsis groups) using EDTA-anticoagulated blood. Left over plasma was used for measuring the plasma von Willebrand factor (vWF) and ADAMTS13 antigen by ELISA assays. The MDW was obtained during routine CBC analysis using a Beckman hematology analyzer. The lactate and high-sensitivity troponin I levels were measured using a Beckman analyzer. Procalcitonin was measured using a Cobas e801 analyzer. Results: The median ISI was twofold higher in the high-probability group than in the low-probability group (p < 0.01) while the median MDW was 34.5% higher in the high-probability group than in the low-probability group (p < 0.01). However, the correlation between the ISI and MDW was only modest (r = 0.66). In addition, significantly higher levels of plasma vWF antigen but lower levels of plasma ADAMTS13 antigen in the high-probability group were found, resulting in significantly higher vWF/ADAMTS13 ratios in the high-probability group than in the low-probability group. Conclusions: The new IntelliSep test along with vWF/ADAMTS13 ratios may be useful for the early diagnosis of sepsis in patients visiting the emergency department, which appears to be superior to the traditional marker, MDW. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20770383
Volume :
13
Issue :
16
Database :
Academic Search Index
Journal :
Journal of Clinical Medicine
Publication Type :
Academic Journal
Accession number :
179382172
Full Text :