Back to Search Start Over

Colorimetric CO2 Detector to Improve Effective Mask Ventilations in Very Preterm Infants: A Pilot Randomized Controlled Study.

Authors :
Kong, Juin Yee
Quek, Bin Huey
Lim, Charis S.E.
Sultana, Rehena
Ng, Yvonne Y.V.
Rajadurai, Victor Samuel
Yeo, Kee Thai
Source :
Neonatology (16617800). 2024, Vol. 121 Issue 4, p494-502. 9p.
Publication Year :
2024

Abstract

Introduction: End-tidal CO2 (ETCO2) detector is currently recommended for confirmation of endotracheal tube placement during neonatal resuscitation. Whether it is feasible to use ETCO2 detectors during mask ventilation to reduce risk of bradycardia and desaturations, which are associated with increased risk of death in preterm babies, is unknown. Methods: This is a pilot randomized controlled trial (NCT04287907) involving newborns 24 + 0/7 to 32 + 0/7 weeks gestation who required mask ventilation at birth. Infants were randomized into groups with or without colorimetric ETCO2 detectors. Combined duration of any bradycardia (<100 bpm) and time below prespecified target oxygen saturation (SpO2) as measured by pulse oximetry were compared. Results: Fifty participants were randomized, 47 with outcomes analysed (2 incomplete data, 1 postnatal diagnosis of trachea-oesophageal fistula). Mean gestational age and birthweight were 28.5 ± 1.9 versus 29.4 ± 1.6 weeks (p = 0.1) and 1,252.7 ± 409.7 g versus 1,334.6 ± 369.1 g (p = 0.5) in the intervention and control arm, respectively. Mean combined duration of bradycardia and desaturation was 276.7 ± 197.7 s (intervention) and 322.7 ± 277.7 s (control) (p = 0.6). Proportion of participants with any bradycardia or desaturation at 5 min were 38.1% (intervention) and 56.5% (control) (p = 0.2). No chest compressions, epinephrine administration, or death occurred in the delivery room. Conclusion: This pilot study demonstrates the feasibility of a trial to evaluate colorimetric ETCO2 detectors during mask ventilation of very preterm infants to reduce bradycardia and low SpO2. Further assessment with a larger population will be required to determine if ETCO2 detector usage at resuscitation reduces risk of adverse outcomes, including death and disability, in very preterm infants. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
16617800
Volume :
121
Issue :
4
Database :
Academic Search Index
Journal :
Neonatology (16617800)
Publication Type :
Academic Journal
Accession number :
179019119
Full Text :
https://doi.org/10.1159/000538083