Back to Search
Start Over
Monitoring TES Loop Gain in Frequency Multiplexed Readout.
- Source :
-
Journal of Low Temperature Physics . Jul2024, Vol. 216 Issue 1/2, p427-435. 9p. - Publication Year :
- 2024
-
Abstract
- We present a method for precise monitoring of the loop gain of transition edge sensors (TES) under electrothermal feedback. The measurement is implemented on the ICE DfMux electronics and operates simultaneously with Digital Active Nulling (DAN). It uses one additional bias sinusoid per TES and does not require any additional readout channels. The loop gain monitor is being implemented on the Simons Array and is an integral part of the baseline calibration strategy for the upcoming LiteBIRD satellite. [ABSTRACT FROM AUTHOR]
- Subjects :
- *MULTIPLEXING
*DETECTORS
*BOLOMETERS
*CALIBRATION
Subjects
Details
- Language :
- English
- ISSN :
- 00222291
- Volume :
- 216
- Issue :
- 1/2
- Database :
- Academic Search Index
- Journal :
- Journal of Low Temperature Physics
- Publication Type :
- Academic Journal
- Accession number :
- 178623307
- Full Text :
- https://doi.org/10.1007/s10909-024-03174-2