Back to Search Start Over

Immunological assay using a solid-state pore with a low limit of detection.

Authors :
Takei, Hiroyasu
Nakada, Tomoko
Leong, lat Wai
Ito, Atsuki
Hanada, Kakeru
Maeda, Hinako
Sohail, Muhammad Shan
Tomiyasu, Kazuhiko
Sakamoto, Osamu
Naono, Norihiko
Taniguchi, Masateru
Source :
Scientific Reports. 7/19/2024, Vol. 14 Issue 1, p1-12. 12p.
Publication Year :
2024

Abstract

Emerging infectious diseases, cancer, and other diseases are quickly tested mainly via immune reactions based on specific molecular recognition between antigens and antibodies. By changing the diameter of solid-state pores, biomolecules of various sizes can be rapidly detected at the single-molecule level. The combination of immunoreactions and solid-state pores paves the way for an efficient testing method with high specificity and sensitivity. The challenge in developing this method is achieving quantitative analysis using solid-state pores. Here, we demonstrate a method with a low limit of detection for testing tumor markers using a combination of immunoreactions and solid-state pore technology. Quantitative analysis of the mixing ratio of two and three beads with different diameters was achieved with an error rate of up to 4.7%. The hybrid solid-state pore and immunoreaction methods with prostate-specific antigen (PSA) and anti-PSA antibody-modified beads achieved a detection limit of 24.9 fM PSA in 30 min. The hybrid solid-state pore and immunoreaction enabled the rapid development of easy-to-use tests with lower limit of detection and greater throughput than commercially available immunoassay for point-of-care testing. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20452322
Volume :
14
Issue :
1
Database :
Academic Search Index
Journal :
Scientific Reports
Publication Type :
Academic Journal
Accession number :
178527196
Full Text :
https://doi.org/10.1038/s41598-024-67112-8