Comparative Analysis of the Neutron–Gamma Discrimination Methods for Scintillators Based on Lithium Glass.

Authors :
Kuzmin, E. S.
Bokuchava, G. D.
Zimin, I. Yu.
Kruglov, A. A.
Kuchinskiy, N. A.
Malyshev, V. L.
Source :
Instruments & Experimental Techniques. Apr2024, Vol. 67 Issue 2, p234-239. 6p.
Publication Year :
2024

Abstract

The characteristics of neutron scintillation detectors built on monolithic and heterogeneous scintillators containing 6Li were studied. The detectors were tested on a thermal neutron beam and on a stand with a source of γ-quanta 60Co. To process the signals received from the detector with a monolithic scintillator, three different γ-radiation discrimination algorithms were used: registration of pulses at a constant threshold and selection according to the pulse shape using two digital signal separation methods: charge integration and pulse duration measurement. For a homogeneous scintillator, pulse shape selection methods work approximately the same when separating thermal neutrons and γ-quanta and are significantly inferior to the method of registration at a constant threshold. In this case, the quality of the n/γ separation is worse than the result obtained with a heterogeneous scintillator when recording at a constant threshold. The purpose of the work is to compare the results of using digital methods for discrimination of gamma-quanta with the results obtained using heterogeneous scintillators. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00204412
Volume :
67
Issue :
2
Database :
Academic Search Index
Journal :
Instruments & Experimental Techniques
Publication Type :
Academic Journal
Accession number :
178316496
Full Text :
https://doi.org/10.1134/S0020441224700465