Back to Search Start Over

Implantation of Hydrophobic Acrylic Toric Intraocular Lens with High-Water Contents Using Swept-Source Optical Coherence Tomography Biometer Integrated with a Surgical Guiding System.

Authors :
Matsumoto, Yukihiro
Azuma, Yuichi
Karasawa, Yasue
Suzuki, Noriyuki
Source :
Clinical Ophthalmology. Apr2024, Vol. 18, p1117-1124. 8p.
Publication Year :
2024

Abstract

Purpose: To evaluate postoperative outcomes after implantation of toric intraocular lenses (IOLs) made of high-water-content hydrophobic acrylic material in Japanese patients using a swept-source optical coherence tomography (SS-OCT) biometer integrated with a surgical guiding system. Patients and Methods: In this prospective observational study, toric IOL models CNW0T3 to CNW0T9 (Alcon) were implanted in 33 eyes of 33 patients and followed-up for one month. Powers and toric models were determined using an SS-OCT biometer ARGOS® Ver 1.5 (Alcon), and the IOLs were aligned using surgical guidance. Differences between planned and actual axis positions at the end of the surgery (misalignment) and rotations from the end of surgery to one month postoperatively were measured. Additionally, postoperative uncorrected visual acuity, refraction, and residual astigmatism were evaluated. Results: Mean and median misalignments were 2.3° (standard deviation [SD]: 1.6, 95% confidence interval [CI]: 1.7– 2.9) and 2°, and those of postoperative rotation were 2.4° (SD: 2.6, 95% CI: 1.5– 3.4) and 2°, respectively. Mean postoperative refraction was 0.06 D (SD: 0.62). Prediction errors within ± 0.5 and ± 1.0 D were 69.7% and 93.9%, respectively. Mean residual astigmatisms were 0.19 D (SD: 0.41), and mean uncorrected visual acuity was 0.00 logMAR (SD: 0.11), and 64% of the eyes scored 20/20 or better. Conclusion: Implantation of high-water-content hydrophobic acrylic toric IOLs using SS-OCT biometry integrated with a surgical guiding system effectively corrected corneal astigmatism with accurate IOL alignment in Japanese patients. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
11775467
Volume :
18
Database :
Academic Search Index
Journal :
Clinical Ophthalmology
Publication Type :
Academic Journal
Accession number :
177004369
Full Text :
https://doi.org/10.2147/OPTH.S456609