Report from the HarmoSter study: different LC-MS/MS androstenedione, DHEAS and testosterone methods compare well; however, unifying calibration is a double-edged sword.

Authors :
Fanelli, Flaminia
Peitzsch, Mirko
Bruce, Stephen
Cantù, Marco
Temchenko, Anastasia
Mezzullo, Marco
Lindner, Johanna M.
Hawley, James M.
Ackermans, Mariette T.
Van den Ouweland, Jody
Koeppl, Daniel
Nardi, Elena
MacKenzie, Finlay
Binz, Pierre-Alain
Rauh, Manfred
Keevil, Brian G.
Vogeser, Michael
Eisenhofer, Graeme
Heijboer, Annemieke C.
Pagotto, Uberto
Source :
Clinical Chemistry & Laboratory Medicine. May2024, Vol. 62 Issue 6, p1080-1091. 12p.
Publication Year :
2024

Abstract

Current liquid chromatography-tandem mass spectrometry (LC-MS/MS) applications for circulating androgen measurements are technically diverse. Previously, variable results have been reported for testosterone. Data are scarce for androstenedione and absent for dehydroepiandrosterone sulfate (DHEAS). We assessed the agreement of androstenedione, DHEAS and testosterone LC-MS/MS measurements among nine European centers and explored benefits of calibration system unification. Androgens were measured twice by laboratory-specific procedures in 78 patient samples and in EQA materials. Results were obtained by in-house and external calibration. Intra- and inter-laboratory performances were valued. Intra-laboratory CVs ranged between 4.2–13.2 % for androstenedione, 1.6–10.8 % for DHEAS, and 4.3–8.7 % and 2.6–7.1 % for female and male testosterone, respectively. Bias and trueness in EQA materials were within ±20 %. Median inter-laboratory CV with in-house vs. external calibration were 12.0 vs. 9.6 % for androstenedione (p<0.001), 7.2 vs. 4.9 % for DHEAS (p<0.001), 6.4 vs. 7.6 % for female testosterone (p<0.001) and 6.8 and 7.4 % for male testosterone (p=0.111). Median bias vs. all laboratory median with in-house and external calibration were −13.3 to 20.5 % and −4.9 to 18.7 % for androstenedione, −10.9 to 4.8 % and −3.4 to 3.5 % for DHEAS, −2.7 to 6.5 % and −11.3 to 6.6 % for testosterone in females, and −7.0 to 8.5 % and −7.5 to 11.8 % for testosterone in males, respectively. Methods showed high intra-laboratory precision but variable bias and trueness. Inter-laboratory agreement was remarkably good. Calibration system unification improved agreement in androstenedione and DHEAS, but not in testosterone measurements. Multiple components, such as commutability of calibrators and EQA materials and internal standard choices, likely contribute to inter-laboratory variability. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
14346621
Volume :
62
Issue :
6
Database :
Academic Search Index
Journal :
Clinical Chemistry & Laboratory Medicine
Publication Type :
Academic Journal
Accession number :
176925986
Full Text :
https://doi.org/10.1515/cclm-2023-1138