Echocardiographic Examination of Atrioventricular and Interventricular Delay Optimization in Cardiac Resynchronization Therapy
- Source :
-
American Journal of Cardiology . May2005, Vol. 95 Issue 9, p1108-1110. 3p. - Publication Year :
- 2005
-
Abstract
- In 21 patients implanted with a biventricular pacing device with programmable interventricular delay (VVd), the myocardial performance index (MPI) was evaluated during spontaneous sinus rhythm, simultaneous biventricular pacing, and sequential biventricular pacing at different VVds and atrioventricular delays (AVds). The AVd-VVd combination associated with the minimum MPI defined patient-tailored biventricular pacing. Simultaneous biventricular pacing significantly improved MPI compared with spontaneous sinus rhythm. An additional improvement was obtained by tailored biventricular pacing. The optimal AVds were significantly shorter during right ventricular preactivation than during left ventricular preactivation. [Copyright &y& Elsevier]
- Subjects :
- *STRESS echocardiography
*HEART diseases
*THERAPEUTICS
*PATIENTS
Subjects
Details
- Language :
- English
- ISSN :
- 00029149
- Volume :
- 95
- Issue :
- 9
- Database :
- Academic Search Index
- Journal :
- American Journal of Cardiology
- Publication Type :
- Academic Journal
- Accession number :
- 17674902
- Full Text :
- https://doi.org/10.1016/j.amjcard.2005.01.028