Back to Search
Start Over
Low‐Power and Fast‐Response Limiting Current‐Type Oxygen Microsensor with a Wide Range Oxygen Concentration.
- Source :
-
IEEJ Transactions on Electrical & Electronic Engineering . May2024, Vol. 19 Issue 5, p796-799. 4p. - Publication Year :
- 2024
-
Abstract
- In this study, we fabricated limiting current‐type oxygen microsensors capable of sensing a wide range of oxygen concentrations. The oxygen microsensors consist of yttria‐stabilized zirconia thin films on spiral Ta2O5 diffusion layers and high‐temperature micro‐hotplates. At temperatures above 575 °C, we observed a clear limiting current that increased proportionally with oxygen concentration between 0.4% and 94%. This wide‐range linearity is attributed to the Knudsen diffusion of the oxygen gas passing through the small gap in the spiral Ta2O5 diffusion layer. © 2023 Institute of Electrical Engineer of Japan and Wiley Periodicals LLC. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 19314973
- Volume :
- 19
- Issue :
- 5
- Database :
- Academic Search Index
- Journal :
- IEEJ Transactions on Electrical & Electronic Engineering
- Publication Type :
- Academic Journal
- Accession number :
- 176635605
- Full Text :
- https://doi.org/10.1002/tee.23939