Low‐Power and Fast‐Response Limiting Current‐Type Oxygen Microsensor with a Wide Range Oxygen Concentration.

Authors :
Akasaka, Shunsuke
Kanno, Isaku
Source :
IEEJ Transactions on Electrical & Electronic Engineering. May2024, Vol. 19 Issue 5, p796-799. 4p.
Publication Year :
2024

Abstract

In this study, we fabricated limiting current‐type oxygen microsensors capable of sensing a wide range of oxygen concentrations. The oxygen microsensors consist of yttria‐stabilized zirconia thin films on spiral Ta2O5 diffusion layers and high‐temperature micro‐hotplates. At temperatures above 575 °C, we observed a clear limiting current that increased proportionally with oxygen concentration between 0.4% and 94%. This wide‐range linearity is attributed to the Knudsen diffusion of the oxygen gas passing through the small gap in the spiral Ta2O5 diffusion layer. © 2023 Institute of Electrical Engineer of Japan and Wiley Periodicals LLC. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
19314973
Volume :
19
Issue :
5
Database :
Academic Search Index
Journal :
IEEJ Transactions on Electrical & Electronic Engineering
Publication Type :
Academic Journal
Accession number :
176635605
Full Text :
https://doi.org/10.1002/tee.23939