Improving Visualization of In-stent Lumen Using Prototype Photon-counting Detector Computed Tomography with High-resolution Plaque Kernel.

Authors :
Yoshinori Funama
Seitaro Oda
Fuyuhiko Teramoto
Yuko Aoki
Isao Takahashi
Shinichi Kojima
Taiga Goto
Kana Tanaka
Masafumi Kidoh
Yasunori Nagayama
Takeshi Nakaura
Toshinori Hirai
Source :
Journal of Medical Physics. Jan-Mar2024, Vol. 49 Issue 1, p127-132. 6p.
Publication Year :
2024

Abstract

The study aimed to compare the performance of photon-counting detector computed tomography (PCD CT) with high-resolution (HR)-plaque kernel with that of the energy-integrating detector CT (EID CT) in terms of the visualization of the lumen size and the in-stent stenotic portion at different coronary vessel angles. The lumen sizes in PCD CT and EID CT images were 2.13 and 1.80 mm at 0°, 2.20 and 1.77 mm at 45°, and 2.27 mm and 1.67 mm at 90°, respectively. The lumen sizes in PCD CT with HR-plaque kernel were wider than those in EID CT. The mean degree of the in-stent stenotic portion at 50% was 69.7% for PCD CT and 90.4% for EID CT. PCD CT images with HR-plaque kernel enable improved visualization of lumen size and accurate measurements of the in-stent stenotic portion compared to conventional EID CT images regardless of the stent direction. [ABSTRACT FROM AUTHOR]