Evaluation of two chemiluminescent immunoassays of ADVIA centaur for hepatitis B serology markers

Authors :
Chen, Dan
Kaplan, Lawrence
Liu, Qiang
Source :
Clinica Chimica Acta. May2005, Vol. 355 Issue 1/2, p41-45. 5p.
Publication Year :
2005

Abstract

Abstract: Background: This study evaluated the performance of 2 Bayer Hepatitis B Chemiluminescent Assays by comparing their performance to the corresponding DPC Immulite 2000 assays. Methods: 953 samples were tested by both Immulite 2000 and Centaur HBsAg methods; and 75 samples were tested by both Immulite 2000 and Centaur anti-HBc IgM methods. Results: The overall agreement between Immulite 2000 and ADVIA Centaur anti-HBc IgM assays was 100%. The agreement between Immulite 2000 and Centaur HBsAg assays was 100% for Immulite 2000 positive samples. The overall agreement for HBsAg negative samples was 99.9% after confirmation result was included. However, discrepancy between the 2 assays was observed in 53 samples which were tested initially as “weakly reactive”(index 1.0–1.95) but were not confirmed subsequently by the neutralization test on Immulite 2000. All 53 samples tested negative by Centaur HBsAg assay at the initial run. Furthermore, one Immulite 2000 HBsAg negative sample was reactive and confirmed by Centaur assay. Conclusions: While the performance of Bayer Hepatitis B HBsAg and aHBc IgM assays are in good agreement with corresponding DPC assays, the Bayer HBsAg assay is more efficient than the Immulite 2000 assay with its better separation of signal and background noise, and its complete automation features. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
00098981
Volume :
355
Issue :
1/2
Database :
Academic Search Index
Journal :
Clinica Chimica Acta
Publication Type :
Academic Journal
Accession number :
17639427
Full Text :
https://doi.org/10.1016/j.cccn.2004.11.031