Back to Search Start Over

The TES-based Cryogenic AntiCoincidence Detector (CryoAC) of ATHENA X-IFU: A Large Area Silicon Microcalorimeter for Background Particles Detection.

Authors :
D'Andrea, M.
Macculi, C.
Lotti, S.
Piro, L.
Argan, A.
Minervini, G.
Torrioli, G.
Chiarello, F.
Ferrari Barusso, L.
Celasco, E.
Gallucci, G.
Gatti, F.
Grosso, D.
Rigano, M.
Brienza, D.
Cavazzuti, E.
Volpe, A.
Source :
Journal of Low Temperature Physics. Feb2024, Vol. 214 Issue 3/4, p164-172. 9p.
Publication Year :
2024

Abstract

We are developing the Cryogenic AntiCoincidence detector (CryoAC) of the ATHENA X-IFU spectrometer. It is a TES-based particle detector aimed to reduce the background of the instrument. Here, we present the result obtained with the last CryoAC single-pixel prototype. It is based on a 1 cm2 silicon absorber sensed by a single 2 mm × 1 mm Ir/Au TES, featuring an on-chip heater for calibration and diagnostic purposes. We have illuminated the sample with 55Fe (6 keV line) and 241Am (60 keV line) radioactive sources, thus studying the detector response and the heater calibration accuracy at low energy. Furthermore, we have operated the sample in combination with a past-generation CryoAC prototype. Here, by analyzing the coincident detections between the two detectors, we have been able to characterize the background spectrum of the laboratory environment and disentangle the primary (i.e. cosmic muons) and secondaries (mostly secondary photons and electrons) signatures in the spectral shape. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00222291
Volume :
214
Issue :
3/4
Database :
Academic Search Index
Journal :
Journal of Low Temperature Physics
Publication Type :
Academic Journal
Accession number :
175634701
Full Text :
https://doi.org/10.1007/s10909-023-03034-5