Back to Search Start Over

Position sensitivity of the TIGRESS 32-fold segmented HPGe clover detector

Authors :
Svensson, C.E.
Hackman, G.
Pearson, C.J.
Schumaker, M.A.
Scraggs, H.C.
Smith, M.B.
Andreoiu, C.
Andreyev, A.
Austin, R.A.E.
Ball, G.C.
Boston, A.J.
Chakrawarthy, R.S.
Churchman, R.
Cowan, N.
Drake, T.E.
Finlay, P.
Garrett, P.E.
Grinyer, G.F.
Hyland, B.
Jones, B.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Mar2005, Vol. 540 Issue 2/3, p348-360. 13p.
Publication Year :
2005

Abstract

Abstract: Sensitivity to the locations of -ray interactions within a 32-fold segmented clover-type HPGe detector has been investigated through Compton scattering of -rays from a collimated source. A mean position sensitivity of 0.44mm at an energy of 373keV is deduced by comparing the average pulse shapes for net charge collecting signals, as well as transient induced signals in neighbouring contacts, from 1007 pairs of three-dimensionally localized interaction points. The reconstruction of individual event locations based on a comparison with the measured set of basis waveforms is presented. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
01689002
Volume :
540
Issue :
2/3
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
17514398
Full Text :
https://doi.org/10.1016/j.nima.2004.11.031