Evaluation of two automated real-time PCR-based quantification methods for whole blood Epstein-Barr viral load.
- Source :
-
Diagnostic Microbiology & Infectious Disease . Feb2024, Vol. 108 Issue 2, pN.PAG-N.PAG. 1p. - Publication Year :
- 2024
-
Abstract
- Quantification of EBV DNA is important in transplantation settings for the diagnosis of post-transplantation. We evaluated the performance of the AltoStar® EBV PCR Kit 1.5 on whole blood specimens: limit of detection, linearity, accuracy, and precision were determined using the WHO NIBSC 09/260 international standard. Results of 69 clinical samples were compared between the AltoStar® EBV PCR Kit 1.5 (altona Diagnostics) and the RealTi m e EBV assay (Abbott). The LoD of the AltoStar® Kit was 148 IU/mL and linearity was between 375 and 500000. A high concordance was found between nominal value of the NIBSC dilutions and the AltoStar EBV result. The total variation ranged from 2.2% to 9.6%. Out of 69 clinical samples tested, there was a high concordance between the 22 paired results within the overlapping linear ranges of both tests. The AltoStar® EBV assay is reliable and accurate for EBV viral load determination on whole blood samples. [ABSTRACT FROM AUTHOR]
- Subjects :
- *VIRAL load
*EPSTEIN-Barr virus
*DETECTION limit
*BLOOD sampling