Back to Search
Start Over
Noise and Signal Analysis of Ir/Au TES With Asymmetrical Slits Parallel to the Electric Current.
- Source :
-
IEEE Transactions on Applied Superconductivity . Jun2005 Part 1 of 3, Vol. 15 Issue 2, p522-525. 4p. - Publication Year :
- 2005
-
Abstract
- A multi-pixel TES array is one of the attractive methods to read-out the incident x-ray position, which simplifies a read-out circuit for imaging applications. We fabricated a ten-pixel Ir/Au TES microcalorimeter with asymmetrical slits parallel to the electric current, and analyzed its noise and signal characteristics. The device was successfully operated in a strong electro-thermal feedback (ETF) mode. However, very large excess noise is observed, which cannot be explained by the noise theory for a single pixel TES. We have modeled the multi-pixel TES array as separated thermal group components, and analyzed noise performance theoretically. As a result, the current noise in our theory is well agreed with the measured noise at a low frequency below 3 kHz. From the x-ray irradiation measurements, ten divided signal groups are observed, which will be associated with the incident x-ray position pixel. The best energy resolution was 18 eV (FWHM) for 5.9 keV x-rays at 0.5 μV bias voltage. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 10518223
- Volume :
- 15
- Issue :
- 2
- Database :
- Academic Search Index
- Journal :
- IEEE Transactions on Applied Superconductivity
- Publication Type :
- Academic Journal
- Accession number :
- 17399772
- Full Text :
- https://doi.org/10.1109/TASC.2005.849896